Guidelines on Generation and Interpretation of Calibration Plots

Your choice of an analytical technique for a particular analysis would be based on its applicability and a conviction that the response will be linear to changes in concentration.

The graphical representation of concentration as x-axis and corresponding instrument response on y- axis is referred to as a calibration plot.

Pic1

Calibration Plot

The concentration of unknown is determined by drawing a perpendicular to the concentration axis from the point of corresponding response. In addition to information on concentration of unknown a calibration plot provides additional information on

Linear Dynamic range which is the concentration range over which the technique gives a linear response to changes in concentration

Minimum detection limit (MDL) which is the minimum amount that can be detected but which cannot be quantified

Minimum detection quantity (MDQ) which is the minimum quantity that can be detected using the technique within the specified level of uncertainty. MDL is always lower than MDQ.

The objective of this article is to familiarize you with the calibration plot, information that you can gain from it and points to be kept in mind before making the calibration plot and its interpretation.

Essential considerations for establishing reliability of calibration plots :

  • A calibration plot should be generated by using the traceable standard of the analyte whose concentration is to be determined.
  • Maintain identical test conditions such as detector wavelength, sample matrix, chromatographic conditions, etc for all calibration point determinations
  • Three points can establish a straight line but calibration plot should be made from at least 5 to 6 points which should include a blank determination.
  • Greater the slope of the calibration plot greater the sensitivity of the method as at lower concentrations the corresponding response signal will be higher than lower slope plots.

Mistakes to be avoided in generation of calibration plots

Some key guidelines are provided which you should verify before proceeding with interpretation and extrapolation of calibration plots

  • A calibration plot should be generated from at least 5 to 6 points and should include a blank determination
Pic2

Too few calibration points

Pic3

Sufficient Calibration Points

  • If a calibration plot is a horizontal straight line it should be rejected straightaway as it would mean that the technique is not responding to changes in concentration of the sample
Pic 4

No relation between concentration and response

Pic3

Correlation between Concentration and Response

    • The calibration line should not be extrapolated beyond the highest concentration recorded as you would not be sure about linearity beyond this point
Pic 6

Invalid Linear Extrapolation

Pic3

Normal Calibration Plot

  • If the highest point is not in a straight line with others do not attempt to draw a curve based on your judgement. In such case you need to have more points to establish the linear relationship beyond the last linear point
Pic 5

Invalid Curvature Extrapolation

Valid Curvature Extrapolation

Valid Curvature Extrapolation

  •  In case of scatter of calibration points use linear regression analysis to establish a best fit line. Do not make personal judgement to decide on the best line which passes through the origin
Pic 8

Personal biased plot

Pic 9

Regression Analysis based Plot

In the present article linear plots have been covered but under real conditions deviations in linear behavior often take place above certain concentrations. The linearity is established on regression analysis and it is quantified using correlation coefficients. These aspects will be covered in a subsequent article.

Please let us know if you have liked the article and do leave your comments.

About Dr. Deepak Bhanot

Dr Deepak Bhanot is a seasoned professional having nearly 30 years expertise beginning from sales and product support of analytical instruments. After completing his graduation and post graduation from Delhi University and IIT Delhi he went on to Loughborough University of Technology, UK for doctorate research in analytical chemistry. His mission is to develop training programs on analytical techniques and share his experiences with broad spectrum of users ranging from professionals engaged in analytical development and research as well as young enthusiasts fresh from academics who wish to embark upon a career in analytical industry.

Comments

  1. Robert Zechariah says:

    Please, always include me in your mailing list.

    Thanks

  2. Kaushik Valambhia says:

    Good and very informative piece of article

  3. i relly learn from this link

  4. I’m greatful about Dr Deepak’s post on the guid and step to ploting calibration graph. Unestly these steps he gave out have served as a guid for me on analytical chemistry which i am studying as a course currently in university, thank you so much Sir.

  5. samba siva says:

    Good analysis…

Speak Your Mind

*

Get Free Access to Continuing Education!

Over 17000 scientists read our free weekly newsletter. Don't be left out!


  • Keep learning to keep succeeding
  • Laboratory Tips, Tricks and Troubleshooting
  • Latest News from the laboratory world
  • Practical and applied knowledge
  • Special discounts and promotions on certificate courses