Useful Tips for Extending the useful life of a GC Capillary Column

Column Oven

GC Capiliary Column mounted inside column oven

Like any other commodity a GC column has a finite life but it is a costly item and should be used with care so that you can take full benefit on your investment. GC column is a delicate item which can retain its operational characteristic s over an extended period provided its handling and operational guidelines are diligently followed.

In this article the possible causes of damage to the column are elaborated upon and guidelines to minimize such damage are suggested.

Breakage

Fused silica capillary columns are strengthened by polyimide coating. However, if scratched or abraded it will break easily. Weakened columns are liable to break due to thermal cycles in the oven or due to vibrations of the oven fan. A broken column can be fixed with a union but the column efficiency begins to degrade if more than two unions are used on a column.

Thermal Damage

Thermal damage becomes significant if a column is subjected for prolonged durations to elevated temperature above its temperature limits and particularly so in in presence of oxygen. At lower temperatures also (above 50°C) a column can get damaged if carrier gas is switched off. It is a good practice to switch off the carrier gas only after cooling the oven to ambient temperature. Common symptoms of thermal damage are peak tailing and column bleed. Gas lines and connections should be periodically inspected for leaks to ensure that oxygen does not gain access to the column.

Chemical Damage

Non-volatile compounds often degrade column performance. Residues resulting from such compounds can be removed by rinsing with solvents such as methanol, methylene chloride or hexane. More than one solvent can also be used in the rinsing process. Starting from an initial temperature of about 40^O to 50^OC, program at a rate of 2 to 3^OC per minute until the upper limit is reached and maintain the temperature for about 1-4 hours.

Inorganic acids or bases should be avoided in the samples. Most of the mineral acids and bases are not very volatile and accumulate in the front of a column. This results in excessive column bleed, peak tailing and overall loss of efficiency. Among organic compounds perfluoroacids acids cause damage to stationary phases. However, chemical damage is generally limited to the front end of the column so trimming ½- 1 m from the front often resolves the problem.

Storage of Columns

After removal from GC the column should be stored in its original box. GC septa should be used to plug the ends to prevent the debris from entering the column. However, before reuse 2 – 4 cm from column end should be trimmed to ensure small pieces of septa from entering the column.

A properly stored column can be used for the years together without any significant drop in its performance efficiency.

About Dr. Deepak Bhanot

Dr Deepak Bhanot is a seasoned professional having nearly 30 years expertise beginning from sales and product support of analytical instruments. After completing his graduation and post graduation from Delhi University and IIT Delhi he went on to Loughborough University of Technology, UK for doctorate research in analytical chemistry. His mission is to develop training programs on analytical techniques and share his experiences with broad spectrum of users ranging from professionals engaged in analytical development and research as well as young enthusiasts fresh from academics who wish to embark upon a career in analytical industry.

Speak Your Mind

*

Get Free Access to Continuing Education!

Over 17000 scientists read our free weekly newsletter. Don't be left out!


  • Keep learning to keep succeeding
  • Laboratory Tips, Tricks and Troubleshooting
  • Latest News from the laboratory world
  • Practical and applied knowledge
  • Special discounts and promotions on certificate courses